Vulnerability database
Every flaw the U.S. government’s cybersecurity agency (CISA) has confirmed attackers are using, plus the new WordPress plugin and theme bugs published each day, sorted by what they mean for a small business. We refresh it twice a day. If you see something your website or office runs, tell us.
6 results
Jun 29, 2020The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuCVE-2020-12047 · Other software · Baxter · reported by ics-cert@hq.dhs.govCritical 9.8
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.
- Product
- Sigma Spectrum Infusion System Firmware
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordus-cert.govus-cert.gov
Jun 29, 2020The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet servCVE-2020-12045 · Other software · Baxter · reported by ics-cert@hq.dhs.govCritical 9.8
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.
- Product
- Sigma Spectrum Infusion System Firmware
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordus-cert.govus-cert.gov
Jun 29, 2020The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operationCVE-2020-12043 · Other software · Baxter · reported by ics-cert@hq.dhs.govCritical 9.8
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted.
- Product
- Sigma Spectrum Infusion System Firmware
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordus-cert.govus-cert.gov
Jun 29, 2020The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits CVE-2020-12041 · Other software · Baxter · reported by ics-cert@hq.dhs.govCritical 9.4
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
- Product
- Sigma Spectrum Infusion System Firmware
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordus-cert.govus-cert.gov
Jun 29, 2020Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unaCVE-2020-12040 · Other software · Baxter · reported by ics-cert@hq.dhs.govCritical 9.8
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
- Product
- Sigma Spectrum Infusion System Firmware
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordus-cert.govus-cert.gov
Jun 29, 2020Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcCVE-2020-12039 · Other software · Baxter · reported by ics-cert@hq.dhs.govLow 2.4
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
- Product
- Sigma Spectrum Infusion System Firmware
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordus-cert.govus-cert.gov
Where this comes from: the exploited list is CISA’s Known Exploited Vulnerabilities catalog. Everything else is the whole National Vulnerability Database, where each record is published by the team that found the bug (Wordfence, Patchstack, WPScan, the vendors themselves and others); we show who. Exploit odds are FIRST’s EPSS, updated daily. Titles and categories are ours, so check the linked record before acting on a detail. For a write-up of the big ones, see our advisories. Every vendor, and every WordPress plugin and theme with a recorded flaw, also has its own page. All of it is free to reuse as JSON, CSV or RSS.
Weekly security brief
The flaws that hit local websites, every Friday.
One short email: what got exploited this week, which WordPress plugins broke, and what to check. Unsubscribe in one click.
Tell us what you're building.
A website, your marketing, a security check, an app, or the thing you've been running out of a spreadsheet for three years. A few lines is plenty.
Not ready to talk yet? Run the free website check and see what Google sees.